PARTNER II: Less oversizing to reduce paravalvular failure with the new generation of SAPIEN 3

The new generation SAPIEN 3 valve was designed to reduce paravalvular aortic regurgitation. The objective of this work was to evaluate the influence of “oversizing” for SAPIEN 3 valve both in relation to the area and the perimeter and determine if calcium ring influences the degree of failure.

A lesser degree of oversizing need it to prevent moderate to severe aortic failure was observed. Being the ideal valve size 0 to 5% more than the size of the ring. The higher the valvular calcification degree index, higher the degree of aortic failure post procedure.

Conclusion

The study showed SAPIEN 3valve has low rate of moderate to severe aortic regurgitation in a low oversizing scenario.

Philipe Blanke
2015-05-20

Original title: Influence of computed tomography-based over-sizing on the incidence of paravalvular regurgitation of a new-generation transcatheter valve: early results of the intermediate-risk S3 cohort PARTNER II trial.

More articles by this author

NONSTEMI: Superiority of early angioplasty in non-ST elevation myocardial infarction

The European guidelines recommend revascularization within 24 hours in high-risk patients with No ST segment elevation myocardial infarction. It is unclear whether these patients...

TOTAL stroke: increased risk of stroke with thromboaspiration

There are many doubts about the benefit of manual thromboaspiration, not only in terms of improvement of the clinical end point but also the...

IN.PACT Global Study: Pharmacological balloon in long lesions of the Superficial Femoral Artery (SFA)

Long lesions (≥ 15 cm), total occlusions, and in-stent restenosis represent a unique treatment challenge in peripheral vascular disease. IN.PACT Global study is a...

Angiographic TOTAL: Sub angiography studio of TOTAL

Myocardial blush is a predictor of mortality after primary angioplasty. Small studies showed a degree of improvement blush with manual thromboaspiration. The present work...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img

Recent Articles

TricValve Transcatheter Bicaval Valve System for Severe Tricuspid Insufficiency: Events at One Year

TricValve is the first bicaval valve system to obtain CE-Mark approval. This device heterotopic device eliminates tricuspid insufficiency reflux into the venous system, reducing...

Long Term Results of the International Chimney Registry

The International Chimney Registry was an observational study aimed at assessing the use of chimney stenting during TAVR either to predict or treat coronary...

CANNULATE TAVR extended study: Impact of Commissural and Coronary Alignment in Coronary Cannulation Following TAVR with Evolut Fx

The new valve Evolut FX has shown better commissural alignment vs. its predecessor Evolut Pro+. Prior studies have already shown commissure alignment facilitates post...