MANTA Closure Device for large-bore arteriotomy closure: Is It Safe?

 Courtesy of Dr. Agustín Vecchia. 

MANTA Closure Device for large-bore arteriotomy closureThe number of transcatheter aortic valve replacements (TAVR) and the placement of aortic endoprosthesis and ventricular assist devices, among others, have grown exponentially worldwide in the last few years. In a short time, many of these procedures will surely become the new standard of care. Logically, this growth comes hand in hand with a progressively more frequent use of large-bore catheters and a higher need for closure devices for the safe performance of these procedures.

 

The following prospective, multicentre study was conducted in three European institutions and was aimed at assessing the safety and efficacy of the MANTA vascular closure device system, designed for 10- to 25-Fr-arteriotomy closure. This device contains a closure unit consisting of polylactic acid (intraarterial toggle), a bovine collagen (extravascular) pad, a nonresorbable polyester suture connecting both, and a stainless steel (nonresorbable) suture lock.

 

The study enrolled 50 patients with a mean age of 79.5 years, who underwent the following:

  • complex coronary angioplasty,
  • ballon aortic valvuloplasty, or
  • transcatheter aortic valve replacement with large-bore catheters (12 to 19 French, i.e., 4 to 6.33 mm).

 

MANTA closure was performed by 9 different operators. Almost the whole group (47 out of 50 patients) underwent TAVR. The mean time to hemostasis with the MANTA device was 2 minutes, 23 seconds.

 

One patient experienced a major vascular complication that required a covered stent and subsequent surgical correction (this patient required emergency surgery for an incorrectly-placed TAVR device). One patient experienced minor bleeding (VARC-2) and one patient experienced a pseudoaneurism. Mortality at 30 days was 8% (n = 4); there were no device-related mortality events. Five patients presented subcutaneous bruising that did not require treatment. The control angiography conducted after the procedure showed patency in the femoral artery in all patients.

 

Conclusion

The authors conclude that the MANTA device demonstrates safety and rapid hemostasis for large-bore arteriotomy closure.

 

Editorial

The importance of this study lies in the fact that most vascular complications observed in large-bore arteriotomies (regardless of the procedure) derive from closure flaws, and MANTA offers the first alternative in that sense.

 

Something to be highlighted is the fact that, although none of the 9 operators in this study had previous experience with the device, this was placed correctly in all patients (in a clinical study context), with acceptable rates of both minor and major complications. This is particularly surprising, taking into account the complication rates observed in recent studies such as “CONTROL” (Closure Device in Transfemoral Aortic Valve Implantation), which enrolled 944 patients undergoing TAVR and compared Prostar XL vs. Perclose Proglide (Abbott Vascular, Redwood City, California). In that case, the rate of major and minor bleeding, according to the VARC classification, was 7.4% and 14.8%, respectively, for patients in the Prostar arm, and 1.9% and 18%, respectively, for patients in the Proglide arm.

 

Courtesy of Dr. Agustín Vecchia. Buenos Aires German Hospital, Argentina.

 

Original title: Percutaneous Plug-Based Arteriotomy Closure Device for Large-Bore Access A Multicenter Prospective Study.

Reference: DOI: 10.1016/j.jcin.2016.12.028 JACC Cardiovascular Interventions.

Authors: Nicolas M. Van Mieghem, MD, PHD, Azeem Latib, MD, Jan van der Heyden, MD, PHD, Lennart van Gils, MD, Joost Daemen, MD, PHD, Todd Sorzano, Jurgen Ligthart, RT, Karin Witberg, CCRN, Thom de Kroon, MD, Nathaniel Maor, Antonio Mangieri, MD, Matteo Montorfano, MD, Peter P. de Jaegere, MD, PHD, Antonio Colombo, MD, Gary Roubin, MD, PHD.


Subscribe to our weekly newsletter

Get the latest scientific articles on interventional cardiology

We are interested in your opinion. Please, leave your comments, thoughts, questions, etc., below. They will be most welcome.

More articles by this author

One-Year Results of ENCIRCLE: Percutaneous Mitral Valve Replacement in Patients Ineligible for Surgery or TEER

Symptomatic mitral regurgitation (MR) in patients who are not candidates for surgery or transcatheter edge-to-edge repair (TEER) remains a highly complex clinical scenario associated...

Can Coronary CT Angiography Replace Invasive Coronary Angiography in Pre-TAVI Coronary Assessment?

Coronary artery disease coexists in approximately half of patients undergoing transcatheter aortic valve implantation, making coronary assessment prior to the procedure essential. Invasive coronary...

Valve-in-Valve in Small Surgical Aortic Bioprostheses: Balloon-Expandable or Self-Expanding? Three-Year Results from the LYTEN Trial

Dysfunction of small surgical aortic bioprostheses represents a challenging scenario for transcatheter aortic valve replacement in the valve-in-valve setting, due to the higher incidence...

Can TAVI Be Safely Performed in Patients With Bicuspid Aortic Valve?

Bicuspid aortic valve (BAV) represents an anatomical challenge for transcatheter aortic valve replacement (TAVR) due to the frequent presence of elliptical annuli, fibroc calcific...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img

Recent Articles

OCT- and IVUS-Guided Coronary Angioplasty in Acute Coronary Syndrome: Long-Term Clinical Outcomes

Percutaneous coronary angioplasty (PCI) in patients with acute coronary syndrome (ACS) has reduced mortality in the acute phase. However, recurrent ACS and target vessel...

One-Year Results of ENCIRCLE: Percutaneous Mitral Valve Replacement in Patients Ineligible for Surgery or TEER

Symptomatic mitral regurgitation (MR) in patients who are not candidates for surgery or transcatheter edge-to-edge repair (TEER) remains a highly complex clinical scenario associated...

Can Coronary CT Angiography Replace Invasive Coronary Angiography in Pre-TAVI Coronary Assessment?

Coronary artery disease coexists in approximately half of patients undergoing transcatheter aortic valve implantation, making coronary assessment prior to the procedure essential. Invasive coronary...