Safety of High Pressure Post-dilation in Bioresorbable Scaffolds

Original Title: Is High Pressure Postdilation Safe in Bioresorbable Vascular Scaffolds? Optical Coherence Tomography Observations After Noncompliant Balloons Inflated at More than 24 Atmospheres.
Reference: Enrico Fabris et al. Catheter Cardiovasc Interv. 2016 Apr;87(5):839-46.

Courtesy of Dr. José Amadeo Guillermo Álvarez.

The mechanical properties of bioresorbable vascular scaffolds (BVS) are different from those of bare metal stents (BMS); given the risk of BVS fracture, high pressure post-dilation is not recommended. However, a thrombosis rate somewhat higher to that of new generation BMS suggests that, in addition to the correct preparation of the plate, it would be necessary some kind of post-dilation to optimize strut apposition and geometry.

The present publication analyzes optical coherence tomography images (OCT) from 22 Absorb bioresorbable scaffolds (Abbott Vascular, Santa Clara California) in 20 lesions from 16 patients undergoing post dilation with non- compliant balloons to 28 ±3.8 atm (range 24 to 40 atm), with a scaffold / balloon diameter/reference vessel diameter ratio of 1.02±0.11. 41% of cases, received high pressure balloons OPN NC (SIS Medical AG, Winterthur, Switzerland) at maximal pressure 30±4.7 atm.

Treated lesions were more complex than those habitually treated with this kind of scaffolds: 35% had moderate to severe calcification (by angiography), 20% involved bifurcation and 55% were ACC/AHA type C.

At final evaluation, there were no fractured devices, there was low residual stenosis area (16±9.6%) with minimal lumen area (OCT) of 5.5±1.4mm2 and only one case of dissected edge (4.5%) that did not require reintervention.

Conclusion
In the context of more complex lesions, post-dilation of bioresorbable scaffolds Absorb with very low compliance balloons and high pressure, with the adequate diameter/ scaffold/ balloon ratio, is safe, since there were no fractured scaffolds and expansion and apposition parameters are comparable to those obtained with less complex lesions.

Editorial Comment
High pressure post dilation has drastically reduced BMS subacute thrombosis rate; however, given the risk of fracture, there are concerns about using this technique with BVS. This is the first study to assess the Absorb BVS integrity after high pressure dilation in vivo and with OCT.

This study analyzed more complex lesions than those habitually treated with these devices, at extremely elevated pressure, with a high percentage of very low compliance balloons.

Outcomes show structure integrity and good area and apposition parameters and, even though further research and clinical follow up is necessary, post dilation could contribute to improve outcomes with this kind of scaffolds.

Courtesy of Dr. José Amadeo Guillermo Álvarez.
Chief of Hemodynamics and Interventional Cardiology at Buenos Aires German Hospital and Buenos Aires British Hospital, Argentina.

More articles by this author

AHA 2024 | SUMMIT

It has been previously shown that the pharmacological treatment of obesity (semaglutide) can reduce cardiovascular events in patients with cardiac failure (CF) and preserved...

AHA 2024 – BPROAD

Hypertension (elevated blood pressure, BP) is the most common comorbidity among diabetic patients and has been associated with higher cardiovascular risk, though as a...

Coronary Artery Disease in Aortic Stenosis: CABG + SAVR vs. TAVR + PCI: Data from Spanish Centers

Multiple randomized studies have shown comparable or superior efficacy of transcatheter aortic valve replacement (TAVR) vs. coronary artery bypass graft (CABG).  However, many of...

Evolution of Small Balloon-Expandable Valves

Small aortic rings (20 mm) have posed a significant challenge for both surgery and transcatheter aortic valve implantation (TAVI) due to their association with an...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img

Recent Articles

Long Term Results of the International Chimney Registry

The International Chimney Registry was an observational study aimed at assessing the use of chimney stenting during TAVR either to predict or treat coronary...

CANNULATE TAVR extended study: Impact of Commissural and Coronary Alignment in Coronary Cannulation Following TAVR with Evolut Fx

The new valve Evolut FX has shown better commissural alignment vs. its predecessor Evolut Pro+. Prior studies have already shown commissure alignment facilitates post...

TRISCEND II: Transcatheter Replacement vs. Medical Treatment for Tricuspid Regurgitation

The EVOQUE device is designed with an intra-annular sealing system that provides excellent anatomical compatibility and an adaptable shape. It is currently available in...