Safety of High Pressure Post-dilation in Bioresorbable Scaffolds

Original Title: Is High Pressure Postdilation Safe in Bioresorbable Vascular Scaffolds? Optical Coherence Tomography Observations After Noncompliant Balloons Inflated at More than 24 Atmospheres.
Reference: Enrico Fabris et al. Catheter Cardiovasc Interv. 2016 Apr;87(5):839-46.

Courtesy of Dr. José Amadeo Guillermo Álvarez.

The mechanical properties of bioresorbable vascular scaffolds (BVS) are different from those of bare metal stents (BMS); given the risk of BVS fracture, high pressure post-dilation is not recommended. However, a thrombosis rate somewhat higher to that of new generation BMS suggests that, in addition to the correct preparation of the plate, it would be necessary some kind of post-dilation to optimize strut apposition and geometry.

The present publication analyzes optical coherence tomography images (OCT) from 22 Absorb bioresorbable scaffolds (Abbott Vascular, Santa Clara California) in 20 lesions from 16 patients undergoing post dilation with non- compliant balloons to 28 ±3.8 atm (range 24 to 40 atm), with a scaffold / balloon diameter/reference vessel diameter ratio of 1.02±0.11. 41% of cases, received high pressure balloons OPN NC (SIS Medical AG, Winterthur, Switzerland) at maximal pressure 30±4.7 atm.

Treated lesions were more complex than those habitually treated with this kind of scaffolds: 35% had moderate to severe calcification (by angiography), 20% involved bifurcation and 55% were ACC/AHA type C.

At final evaluation, there were no fractured devices, there was low residual stenosis area (16±9.6%) with minimal lumen area (OCT) of 5.5±1.4mm2 and only one case of dissected edge (4.5%) that did not require reintervention.

Conclusion
In the context of more complex lesions, post-dilation of bioresorbable scaffolds Absorb with very low compliance balloons and high pressure, with the adequate diameter/ scaffold/ balloon ratio, is safe, since there were no fractured scaffolds and expansion and apposition parameters are comparable to those obtained with less complex lesions.

Editorial Comment
High pressure post dilation has drastically reduced BMS subacute thrombosis rate; however, given the risk of fracture, there are concerns about using this technique with BVS. This is the first study to assess the Absorb BVS integrity after high pressure dilation in vivo and with OCT.

This study analyzed more complex lesions than those habitually treated with these devices, at extremely elevated pressure, with a high percentage of very low compliance balloons.

Outcomes show structure integrity and good area and apposition parameters and, even though further research and clinical follow up is necessary, post dilation could contribute to improve outcomes with this kind of scaffolds.

Courtesy of Dr. José Amadeo Guillermo Álvarez.
Chief of Hemodynamics and Interventional Cardiology at Buenos Aires German Hospital and Buenos Aires British Hospital, Argentina.

More articles by this author

Is IVUS Always Necessary for Left Main Coronary Artery PCI?

Percutaneous coronary intervention (PCI) of the unprotected left main coronary artery is a highly complex procedure because of the large amount of myocardium at...

Dual-Prep Registry: Atherectomy and IVL for Severe Coronary Calcification

Severe coronary calcification remains one of the most challenging scenarios in percutaneous coronary intervention (PCI). Although rotational or orbital atherectomy and intravascular lithotripsy (IVL)...

Prehospital heparin in STEMI: A safe strategy associated with improved early reperfusion

Early reperfusion remains the main prognostic determinant in patients with ST-segment elevation myocardial infarction (STEMI). Although primary percutaneous coronary intervention (PCI) is the treatment...

Plaque Ruptures in Non-Culprit Arteries: Follow-Up With Intravascular Imaging

Plaque rupture remains one of the most important pathophysiological mechanisms in acute coronary syndromes. However, not all ruptures manifest clinically as ischemia, myocardial infarction,...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img
Jornadas Guatemala 2026

Recent Articles

Coronary artery disease progression after transcatheter aortic valve replacement: quantitative coronary angiography and Quantitative Flow Ratio analysis

The coexistence of coronary artery disease and severe aortic stenosis is common among patients undergoing transcatheter aortic valve replacement (TAVR). However, the impact of...

UNICORN Technique to Prevent Coronary Obstruction During TAVI: Initial Results From a Multicenter Study

Coronary obstruction is an uncommon but potentially catastrophic complication of transcatheter aortic valve implantation (TAVI), particularly in valve-in-valve procedures, TAV-in-TAV interventions, or in patients...

Supera vs. Eluvia at 3 Years in Severely Calcified Femoropopliteal Lesions

Severe calcification remains one of the main predictors of restenosis and the need for repeat revascularization following endovascular treatment of femoropopliteal disease. In this...