BuMA stent: new sirolimus-eluting stent with bio-degradable polymer

Uncovered struts compromise re-endothelization and may induce thrombosis. The new stent approved in China and already used in more than 50,000 patients, is a sirolimus-eluting device with a bio-degradable polymer. Animal studies showed rapid and complete endothelialization. The study included 20 patients with stable angina randomized on a 1:1 ratio to receive, in the same lesion and with an overlap, the proximal sirolimus-eluting stent and the distal everolimus-eluting stent vs the proximal everolimus-eluting stent and the distal sirolimus-eluting stent. The primary endpoint was the coverage of struts after three months and the secondary endpoint was the anual coverage and the clinical events. 

After 3 months, there was no difference in relation to the coverage of struts and intimal thickness between the two stents. There was a difference was observed in luminal loss (everolimus 11.3 ± 4.1 mm vs. sirolimus 10.3 ± 3.8 mm, p<0.001), poor apposition (everolimus 0.4% vs. sirolimus 0.1%, p<0.001). These results were maintained after 12 months. The struts of the stents were found well covered after 3 and 12 months, with a difference in favor of the sirolimus-eluting stent (99.2% vs 98.2%, p<0.001). 

Conclusion: the BuMA sirolimus-eluting stent presents an earlier endothelization than the everolimus-eluting stent.

jingbo_hou_europcr
Jingbo Hou
2013-05-22

Título original: A prospective randomised controlled 3- and 12-month OCT study to evaluate the endotelial healing between a novel sirolimus-eluting stent BuMA and an everolimus-eluting stent XIENCE V.

More articles by this author

REPRISE II: results obtained with the new valve according to VARC criteria

The objective was to evaluate the safety and efficacy of the Lotus Valve System to treat patients with severe aortic stenosis and high surgical...

DEFLECT I: cerebral protection device in TAVI

Strokes after TAVI occur in between 3.8 and 6.7% of the cases, with new MRI lesions being almost the rule. The TriGuard EDD device...

DISCOVER: new low profile valve with little post-implant regurgitation

The objective of the study was to assess events after 30 days with the DirectFLOW Medical Valve device in patients with severe aortic stenosis...

JUPITER: transapical TAVI device

We evaluated 88 consecutive patients (at this phase) with severe aortic stenosis who received the second-generation device for transapical access. Survival at thirty days...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img

Recent Articles

Long-Term Predictors of Valve Deterioration After TAVI

Transcatheter aortic valve implantation (TAVI) has become an innovative technology for the treatment of patients with severe aortic valve stenosis. Initially implemented in elderly...

Coronary access after TAVI with fourth- and fifth-generation Evolut valves: the EPROMPT-CA study

The expansion of transcatheter aortic valve implantation (TAVI) toward younger and lower-risk patients has increased the relevance of post-procedural coronary access. In this context,...

Morpheus Global Registry: Safety and efficacy of the long tapered BioMime™ Morph stent in complex coronary lesions

Percutaneous coronary intervention in long coronary lesions continues to represent a technical and clinical challenge, in which the use of conventional cylindrical stents may...