Aortic Angulation: Impact on the Success of TAVR

Original Title: Aortic Angulation Attenuates Procedural Success Following Self-Expandable But Not Balloon-Expandable TAVR.

Reference: Abramowitz Y  et al. JACC Cardiovasc Imaging. 2016 Jul 8. [Epub ahead of print].

 

Impact of Post TAVR Aortic Angulation

The angulation degree between aorta and heart can make precise valve implantation challenging, especially extreme angulation, such as a horizontal aortic root.

As obvious as it may seem, there is limited data on the impact of aortic angulation in TAVR.

The aim of this study was to assess the impact of aortic angulation immediately after transcatheter aortic valve replacement (TAVR).

The study included 582 patients undergoing TAVR that had prior procedure CT scans available for aortic angulation evaluation.

End points were considered according to the Valve Academic Research Consortium-2 definitions.

Mean aortic angulation was 47.3 ± 8.7°, and patients were divided in two groups: <48° or ≥48° angulation.

In 480 patients receiving balloon expandable valves, the aorta did not influence acute periprocedural success or short term clinical outcomes.

On the contrary, in 102 patients receiving self-expandable valves, increased aortic angulation attenuated acute periprocedural success (area under the curve: 0.73; CI 95% 0.61 to 0.85; p = 0.008).

The numerical cutoff for aortic angulation with the highest sum of sensitivity and specificity for device success was in fact ≥48° (sensitivity 85%, specificity 61%).

Patients with ≥48 angulation receiving self-expandable valves often needed a second valve and post dilation, had longer fluoroscopy time and increased device embolization and mild or greater than mild paravalvular regurgitation.

Mortality at 30 days was not influenced by device angulation.

 

Conclusion

Increased aortic root angulation can have a negative impact in acute periprocedural success in patients receiving self-expandable aortic valves but no in patients receiving balloon expandable valves.

 

Editorial Comment

Future studies with new devices should include aortic angulation as a factor to assess. New self-expandable repositionable and retrievable valves could solve this problem: during release (when presenting valve waist at annulus level) it is possible to push the whole system so that it “copies” the angulation between aorta and heart and to complete a practically horizontal release.

 

We value your opinion. You are more than welcome to leave your comments, thoughts, questions or suggestions here below.

More articles by this author

Hemodynamic Outcomes of Edge-to-Edge Repair in Degenerative and Functional Mitral Regurgitation

Transcatheter edge-to-edge mitral valve repair (M-TEER) has become an established therapeutic option for mitral valve disease. Among the available techniques, M-TEER using the MitraClip...

SAPIEN 3 TAVI Durability: Ten-Year Follow-Up in Intermediate-Risk Patients

The durability of transcatheter bioprosthetic valves used in TAVI remains one of the key unanswered questions as indications continue to expand toward patients with...

Inflammation after TAVI: An Emerging Therapeutic Target?

Conduction disturbances and the need for permanent pacemaker implantation remain common complications following TAVI, with an incidence approaching 15%. Although they have traditionally been...

Coronary Obstruction During TAVI: A New Volumetric Index to Consider

Coronary obstruction during TAVI is an uncommon but potentially catastrophic complication, particularly in valve-in-valve procedures, in anatomies with small sinuses of Valsalva, low coronary...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img
Jornadas Guatemala 2026

Recent Articles

Is IVUS Always Necessary for Left Main Coronary Artery PCI?

Percutaneous coronary intervention (PCI) of the unprotected left main coronary artery is a highly complex procedure because of the large amount of myocardium at...

Dual-Prep Registry: Atherectomy and IVL for Severe Coronary Calcification

Severe coronary calcification remains one of the most challenging scenarios in percutaneous coronary intervention (PCI). Although rotational or orbital atherectomy and intravascular lithotripsy (IVL)...

Hemodynamic Outcomes of Edge-to-Edge Repair in Degenerative and Functional Mitral Regurgitation

Transcatheter edge-to-edge mitral valve repair (M-TEER) has become an established therapeutic option for mitral valve disease. Among the available techniques, M-TEER using the MitraClip...