After a transcatheter aortic valve replacement (TAVR), the incidence of permanent pacemaker (PPM) implantation ranges from 6% to 30% in different studies, depending on the valve type used. This is due to trauma, ischemia, haemorrhage, and atrioventricular node and infranodal tissue oedema, provoking ventricular dyssynchrony similar to that caused by complete left bundle branch block. The evolution of patients requiring permanent pacemaker after TAVR has not been fully studied.

 

This study included 9785 patients from 229 hospitals; 651 (6.65%) of these subjects received a permanent pacemaker.

 

Patients receiving a permanent pacemaker were more frequently men with a higher STS predicted risk of mortality score (7.3% vs. 6.7%; p = 0.004), who received aortic valvuloplasty less often and presented more conduction defects in the baseline echocardiogram.

 

During the procedure, patients requiring a permanent pacemaker received larger valves, with higher oversizing. Mean time between valve replacement and permanent pacemaker implantation was three days.

 

Self-expanding valves (CoreValve) required the implantation of permanent pacemakers more frequently than balloon-expandable valves (E. Sapiens). Extreme-risk or inoperable patients required a permanent pacemaker more frequently than high-risk patients. Hospital stays were longer for patients requiring a PPM.

 

At a year, the need for a permanent pacemaker was associated with an increase in mortality (24.1% vs. 19.6%; hazard ratio [HR]: 1.31; 95% confidence interval [CI]: 1.09 to 1.58) and an increase in a composite of mortality and heart failure (37.3% vs. 28.5%; HR: 1.33; 95% CI: 1.13 to 1.56).

 

The following were predictors of the need for permanent pacemaker implantation after TAVR at 30 days:

  • age;
  • prior conduction defect;
  • aortic valve area ≤0.75 cm2;
  • use of self-expanding valves.

 

Conclusion

The need for a pacemaker is a common complication following TAVR, and it is associated with higher mortality and hospitalization due to heart failure at one year.

 

Editorial

Prior studies argued that permanent pacemaker implantation did not affect outcomes. This study, however, proves the opposite: it is associated to a hard event, mortality, and a higher need for hospitalization due to heart failure. This also increases healthcare costs.

 

The upcoming challenge lies in the development of devices that lower the need for a permanent pacemaker while maintaining or improving current performance.

 

Courtesy of Dr. Carlos Fava. Buenos Aires Favaloro Foundation, Argentina.

 

Original title: Incidence, Predictors, and Outcomes of Permanent Pacemaker Implantation Following Transcatheter Aortic Valve Replacement. Analysis from the U.S. Society of Thoracic Surgeons/American College of Cardiology TVT Registry.

Reference: Opeyemi O. FadaHunsi, et al. J Am Coll Cardiol Intv 2016;9:2189-99.


Subscribe to our weekly newsletter

Get the latest scientific articles on interventional cardiology

We are interested in your opinion. Please, leave your comments, thoughts, questions, etc., below. They will be most welcome.

More articles by this author

VECTOR: First Percutaneous Aorto-Coronary Bypass Case, a New Conceptual Approach

Coronary obstruction represents one of the most severe complications associated with transcatheter aortic valve implantation, particularly in valve-in-valve scenarios involving surgical bioprostheses, narrow aortic...

Is it safe to use negative chronotropic drugs early after TAVI?

TAVI is associated with a relevant incidence of conduction system disturbances and the development of atrioventricular block that may require permanent pacemaker implantation. Many...

Transapical TMVR in High Risk Patients: Intrepid 5-Year Outcomes

Moderate to severe mitral valve regurgitation (MR) continues is still a high prevalence condition with bad prognosis, particularly among the elderly with left ventricular...

Impact of Balloon Post-Dilation on the Long-Term Durability of Bioprostheses after TAVR

Balloon post-dilation (BPD) during transcatheter aortic valve replacement (TAVR) allows for the optimization of prosthesis expansion and the reduction of residual paravalvular aortic regurgitation....

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img

Recent Articles

VECTOR: First Percutaneous Aorto-Coronary Bypass Case, a New Conceptual Approach

Coronary obstruction represents one of the most severe complications associated with transcatheter aortic valve implantation, particularly in valve-in-valve scenarios involving surgical bioprostheses, narrow aortic...

Comparison of strategies: NMA of IVUS, OCT, or angiography in complex lesions

Percutaneous coronary intervention (PCI) in complex lesions continues to represent a technical challenge in contemporary interventional cardiology. Angiography, although it remains the most widely...

Is upper-limb aerobic training an effective alternative to lower-limb exercise in peripheral artery disease?

Peripheral artery disease is associated with impaired functional capacity, reduced walking distance, and poorer quality of life, and structured exercise is a class I...