REPRISE III: Need for Pacemaker and Paravalvular Leak with Lotus

A recent presentation at London Valves took a closer look at the REPRISE III trial as regards two points that generated much discussion, such as need for a pacemaker and paravalvular leak.

REPRISE III: Leak paravalvular y necesidad de marcapasos con Lotus

REPRISE III had previously demonstrated that the Lotus is noninferior to first-generation CoreValve and Evolut R in terms of safety. Additionally, it proved to be superior for the primary efficacy combined endpoint at 1 year (death, disabling stroke, and moderate or greater paravalvular leak).


Read also: Impella Improves Safety in High Risk Unprotected Left Main PCI”.


The definition of the REPRISE III composite safety endpoint did not include need for a pacemaker, which was stuck at 35.5% with Lotus vs. 19.6% for the combined both-generation CoreValve group (p < 0.001).

 

On the other hand, the composite efficacy endpoint was basically driven by the reduction in moderate/severe leaks (2% vs. 11.1%; p < 0.001).

 

The Lotus valve is specifically designed for highly calcified disease anatomies, thanks to its repositionability and its small cell frame that conforms to rigid calcium structure. On the other hand, its adaptive urethane seal conforms to irregular surfaces.


Read also: Small Annulus: Should We Start Choosing the Valve?”


Patients were categorized by calcium load observed in the computerized tomography scan across three tertiles (calcium in the leaflets, annulus, and outflow tract); ≥795.3 mm3 were considered as severe calcification. A total of 98 out of 297 CoreValve patients presented severe calcification, and so did 195 out of 577 patients in the Lotus valve group.

 

At 1 year, moderate/severe paravalvular leak was significantly greater among patients who received the CoreValve device and presented severe calcification vs. Lotus patients. This difference was not observed in the two tertiles with lower calcium burden.

 

These data might suggest that transcatheter aortic valve replacement (TAVR) with the Lotus valve may be more beneficial in treating patients with severe calcification than other valves.

 

Most patients who required a pacemaker after Lotus implantation needed it due to third-degree atrioventricular block or Mobitz Type 2, and this did not affect clinical results or patient quality of life.

 

Implantation depth was a key predictor of need for a pacemaker, and perfecting the Lotus implantation technique reduced the overall number of pacemakers required. However, this difference was not statistically significant.

 

A 30% rate of need for a pacemaker with Lotus does not seem to be acceptable now that TAVR is moving towards intermediate- and low-risk populations.

 

The end of this story will be told by the new Lotus Edge, which incorporates a depth guard system that might lower the pacemaker implantation rate from 30% to 10%.

 

Original title: The Impact of Calcium Burden on Clinical Outcomes and Paravalvular Regurgitation in the REPRISE III Randomized Controlled Clinical Trial.

Reference: Makkar R et al. Presented at: PCR London Valves 2017. September 25, 2017. London, England.


Subscribe to our weekly newsletter

Get the latest scientific articles on interventional cardiology

We are interested in your opinion. Please, leave your comments, thoughts, questions, etc., below. They will be most welcome.

More articles by this author

Coronary revascularization before TAVI: prior PCI or conservative management?

The coexistence of coronary artery disease (CAD) in patients with severe aortic stenosis undergoing TAVI is common, with a reported prevalence ranging from 30%...

Percutaneous Mechanical Aspiration versus Surgical Treatment of Tricuspid Valve Endocarditis

Tricuspid valve infective endocarditis (TVIE) accounts for approximately 5% to 10% of all cases of infective endocarditis. Surgical treatment remains the standard therapy in...

CRT 2026 | TAVI-CLOSE Trial: Dual Suture vs Suture + Plug for Vascular Closure After Transfemoral TAVI

The transfemoral approach is the predominant strategy for transcatheter aortic valve implantation (TAVI). Although vascular complications are currently less frequent, they remain relevant determinants...

CRT 2026 | NAVITOR IDE: Hemodynamic Outcomes and 5-Year Durability of an Intra-Annular Self-Expanding Transcatheter Aortic Valve

As TAVI expands into younger populations and patients with lower surgical risk, prosthesis durability has become a key aspect of long-term management. The NAVITOR...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img

Recent Articles

Drugs for the Treatment of No-Reflow During PCI

The no-reflow phenomenon is one of the most frustrating complications of primary angioplasty (pPCI), reflecting persistent microvascular damage that, in the mid- to long-term,...

Coronary revascularization before TAVI: prior PCI or conservative management?

The coexistence of coronary artery disease (CAD) in patients with severe aortic stenosis undergoing TAVI is common, with a reported prevalence ranging from 30%...

Rotational atherectomy and its technical secrets: use of floppy or ES guidewire

Rotational atherectomy (RA) remains a very useful tool in the management of severe coronary calcification. However, many of its technical aspects rely more on...