Longitudinal Deformation of a Stent with the POT Technique

The fractal geometry of coronary bifurcations offers a mismatch between the proximal diameter of the main branch and the distal branches that make up the bifurcation. Angioplasty in this area can be performed through different techniques depending on plaque distribution, and the particular geometry of each bifurcation.

Deformación longitudinal de un stent con la técnica de POT

Overall, and regardless of the strategy used, there is a common step: the proximal optimization technique (POT). Since the stent diameter chosen must follow distal vessel caliber (to prevent vascular trauma), the proximal segment may often be under-expanded. If the mismatch is significant, successive POTs may result in stent overexpansion, leading to structural deformation. There is not much literature on this “longitudinal behavior.”

The aim of this paper was to analyze and describe the longitudinal patterns of stent deformation during POT overexpansion, both on a test bench and in the clinical analysis of a retrospective registry.

In evaluating the behavior on the test bench (siliconized tube), the POT was simulated in two different models with a common proximal main segment (internal diameter >6.0 mm) and a 1-mm long transition connecting to the distal segment (3.0-mm or 3.5-mm platforms).

Testing was performed with 5 different stent types: Orsiro, Resolute Onyx, Sinergy, Sinergy Megatron, Ultimaster, and Xience (all of them 28-mm long), with an evaluation of different lengths over the proximal area. Optical coherence tomography (OCT) was performed systematically twice after stent implantation, and with every repeat POT.

Read also: Coronary Physiology Is Useful in Chronic Kidney Disease.

For clinical control, researchers analyzed the OCT database of bifurcation lesion angioplasties from four participating centers over 2 years.

In bench testing, 15 tests were performed with the 3.0-mm platform, and 14 tests with the 3.5-mm platform. When testing the 3.0 mm platform, significant elongation was observed with each 0.5-mm increase in POT, with a maximum elongation of 4.31 ± 1.47 mm when reaching an overexpansion of 5.5 mm.

The same effect was evidenced in the 3.5-mm platform with a significant elongation every 0.5 mm, with a maximum elongation of 2.87 ± 0.94 mm at the proximal level (meaning a relative elongation of 19.45%).

Read also: Progress-CTO Score: A Key New Tool to Plan CTO.

A significant difference was observed when comparing mean elongation in different proximal lengths (10 mm: 1.34 ± 1.36 mm; 15 mm: 1.64 ± 1.30 mm; 20 mm: 2.12 ± 1.89 mm; p = 0.05). When analyzing the different stent types, no significant differences were observed, with no stent fractures in the OCT analysis.

Clinical analysis included 36 cases (most in the distal main coronary artery). In this sense, OCT imaging showed a mean elongation of 2.22 ± 1.35 mm (p < 0.01).

Conclusions

When analyzing a test bank and bifurcation cases from an OCT registry, a relevant elongation at the proximal level was evidenced in the simulated model with each 0.5 mm overexpansion. These data were confirmed in the retrospective analysis of the OCT registry of bifurcation lesions, which showed an elongation greater than 2 mm in the proximal segment. These results should be taken into account when deciding on the stent, mainly in segments such as the left main coronary artery, since poor apposition could result in vessel revascularization failure with the subsequent clinical consequences that this entails.

Dr. Omar Tupayachi

Dr. Omar Tupayachi.
Member of the Editorial Board of SOLACI.org.

Original Title: Bench test and in vivo evaluation of longitudinal stent deformation during proximal optimization.

Font: EuroIntervention 2022;18:83-90. DOI: 10.4244/EIJ-D-21-00824.


Subscribe to our weekly newsletter

Get the latest scientific articles on interventional cardiology

More articles by this author

ACC 2026 | DKCRUSH VIII: IVUS or angiography to guide PCI in complex coronary bifurcations

Intracoronary imaging guidance has become an established recommended strategy in complex coronary lesions. In the specific setting of complex bifurcations, uncertainty remained regarding the...

ACC 2026 | OPTIMAL: IVUS Guidance in PCI of the Unprotected Left Main Coronary Artery

Percutaneous coronary intervention (PCI) is considered an equivalent alternative to coronary artery bypass surgery in patients with left main coronary artery (LMCA) stenosis and...

ACC 2026 | IVUS-CHIP Trial: Intravascular ultrasound–guided versus angiography-guided complex PCI

Optimization of percutaneous coronary intervention (PCI) in complex lesions remains a relevant clinical challenge. In this context, the IVUS-CHIP trial was designed to evaluate...

ACC 2026 | ALL-RISE Trial: Coronary Physiological Assessment Using FFRangio

Coronary physiological assessment using pressure-wire techniques (FFR/iFR) carries a Class IA recommendation in ACC/AHA guidelines; however, its use remains limited due to factors such...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img

Recent Articles

Experience with the intra-annular self-expanding Navitor valve: data from the STS/ACC TVT registry

The expansion of TAVI, with the introduction of new-generation devices, has prioritized not only periprocedural safety, but also the preservation of coronary access, more...

Therapeutic strategies in carotid free-floating thrombus: evidence and controversies

Carotid free-floating thrombus (cFFT) is a rare entity with a high embolic risk, associated with acute neurological events such as stroke or transient ischemic...

The Two Sides of the Coin: What Do CHAMPION-AF and CLOSURE-AF Teach Us About Left Atrial Appendage Closure?

Letter to the editor: Juan Manuel Pérez Asorey Percutaneous left atrial appendage closure (LAAO) is currently going through one of the most interesting stages of...