SAFE-TAVI Study: Safety and Efficacy of the Pressure Sensor and Pacing Guidewire

Currently, we observe an increase in the number of transcatheter aortic valve replacement (TAVR) procedures due to its expansion toward a younger population with lower risk. For this reason, it is crucial to maintain continuous technological innovation to minimize post-procedural complications.

Estudio SAFE-TAVI: Seguridad y eficacia de la guía para marcapasear y censar presiones

The FDA has approved the SavyWire, a preformed 0.035-caliber guidewire that offers three essential functions: 1) it facilitates prosthesis transport and positioning; 2) it allows for continuous hemodynamic monitoring through a distal optical fiber sensor; 3) it enables pacing through the left ventricle.

The main objective of this prospective multicenter study, named SAFE-TAVI (SavvyWire Efficacy and Safety in Transcatheter Aortic Valve Implantation Procedures), was to assess the safety and efficacy of the SavyWire in TAVR, using both balloon-expandable and self-expandable valves.

The primary endpoint (PEP) was defined as rapid and effective pacing with a significant decrease in pressure (reduction of systolic pressure below 60 mmHg). The secondary safety endpoint (SSE) included the absence of major guidewire-related complications.

Researchers analyzed a total of 119 patients with a mean age of 82.2 ± 5.9 years, 49.6% of whom were women. The average STS Score was 3.85. All patients were treated via transfemoral access. Balloon-expandable valves were used in 37.8% of cases. Predilation was performed in 74.8% of subjects and postdilation in 11.8% of patients.

Read also: Heterotopic Tricuspid Intervention: TricValve One-Year Outcomes.

Regarding the PEP, a drop in systolic pressure below 60 mmHg was achieved in 98.3% of patients, and the mean arterial systolic pressure during rapid pacing was 46.6 ± 11.3 mm Hg. The SSE was reached in 99.2% of patients, with no mortality, stroke, guidewire kinking, or ventricular perforation events.

Conclusion

The results of this study demonstrate the safety and efficacy of the SavyWire during TAVR. This device can contribute to minimizing intervention during the procedure and improving clinical outcomes after transcatheter aortic valve implantation.

Dr. Andrés Rodríguez.
Member of the Editorial Board of SOLACI.org.

Original Title: Safety and Efficacy of Transcatheter Aortic Valve Replacement With a Pressure Sensor and Pacing Guidewire SAFE-TAVI Trial.

Reference: Ander Regueiro, MD, PHD et al J Am Coll Cardiol Intv 2023.


Subscribe to our weekly newsletter

Get the latest scientific articles on interventional cardiology

More articles by this author

Is it really necessary to monitor all patients after TAVR?

Conduction disorders (CD) after transcatheter aortic valve replacement (TAVR) are a frequent complication and may lead to the need for permanent pacemaker implantation (PPI)....

Is it really necessary to monitor all patients after TAVR?

Conduction disorders (CD) after transcatheter aortic valve replacement (TAVR) are a frequent complication and may lead to the need for permanent pacemaker implantation (PPI)....

Transradial Aortic Valvuloplasty: Is Minimalism Worth It?

Balloon aortic valvuloplasty (BAV) has historically been used either as a bridge strategy, an assessment tool, or even palliative treatment in severe aortic stenosis...

Atrial Fibrillation After Percutaneous Patent Foramen Ovale Closure: Cohort Study with Continuous Implantable Cardiac Monitoring

Atrial fibrillation (AF) is a recognized complication following percutaneous closure of a patent foramen ovale (PFO), with reported incidences of up to 30% during...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img

Recent Articles

Coronary Perforations and Use of Covered Stents: Safe and Effective Long-Term Strategy?

Coronary perforations remain one of the most serious complications of percutaneous coronary intervention (PCI), especially in cases of Ellis ruptures type III. In these...

Is it really necessary to monitor all patients after TAVR?

Conduction disorders (CD) after transcatheter aortic valve replacement (TAVR) are a frequent complication and may lead to the need for permanent pacemaker implantation (PPI)....

Is it really necessary to monitor all patients after TAVR?

Conduction disorders (CD) after transcatheter aortic valve replacement (TAVR) are a frequent complication and may lead to the need for permanent pacemaker implantation (PPI)....