TAVI in Horizontal Aorta: Neo2 vs. Sapien Ultra

There are controversies regarding transcatheter aortic valve implantation (TAVI) in cases of challenging anatomical complexity, such as horizontal aorta (HA). This is primarily because aortic angulation (AA) impacts the procedure differently depending on the valve type used. As previously studied, angulation complicates crossing, implantation, and coaxial positioning with commissural alignment.

angulación aórtica post TAVR

In the study presented by Eckel C. et al., the clinical outcomes of self-expanding ACURATE Neo2 valve and balloon-expandable Edwards SAPIEN Ultra valve were compared using propensity score matching. The study retrospectively included consecutive patients treated with transfemoral TAVI from January 2017 to January 2023 using Ultra and Neo2 systems at two high-volume centers in Germany.

AA was measured as the angle between the virtual basal ring of the aortic annulus and the horizontal plane. Horizontal AA was defined as an angle greater than 48°, and severe AA was defined as an angle that exceeded 51.7° (measured by 3Mensio on the coplanar plane).

The primary endpoint (PEP) was in-hospital mortality, need for bailout maneuvers during the procedure, and technical success. Secondary endpoints included 30-day mortality, 30-day stroke, 30-day device success, and procedural complications.

Read also: Complex PCI in Octogenarian.

The final cohort comprised 492 patients (Neo2 n = 246; S3 Ultra n = 246), with an average age of 81.9 years; 43.9% of subjects were women. The Neo2 group had a higher rate of predilation (18.0% vs. 93.9%, p < 0.001) and postdilation (6.5% vs. 33.7%, p < 0.001), with high technical success in both cases (93.1%). When evaluating device success, Ultra valves (80.5% vs. 88.6%, p = 0.018) had a significantly lower performance rate, mainly due to a higher incidence of elevated gradients (≥20 mmHg; 11.9% vs. 1.6%, p < 0.001). Device migration and embolization rates (0.4% vs. 2.0%, p = 0.216), as well as bailout maneuvers (0.8% vs. 0.4%, p = 1.000), were comparable between devices.

There were no statistical differences regarding all-cause mortality (1.3% vs. 2.7%, p = 0.330). However, an analysis of hemodynamic variables related to long-term valve performance showed hemodynamic superiority favoring Neo2, with a lower incidence of patient-prosthesis mismatch (PPM) (12.0% vs. 3.3%, p = 0.001) and lower gradients.

Regarding paravalvular leak (PVL), balloon-expandable valves showed a lower incidence of at least moderate PVL (0.0% vs. 3.3%, p = 0.007).

Read also: High Gradients After Valve-in-Valve.

When stratifying angulation as mild (20-43°), moderate (43-50.3°), severe (50.4-56.1°), or extreme (56.2-87°), implant depth was shallower in severe and extreme angulations (5.0 mm vs. 5.0 mm vs. 4.0 mm vs. 4.0 mm, p = 0.031), with higher numerical PVL (1.6% vs. 0.7% vs. 3.1% vs. 1.5%, p = 0.070).

Conclusions

Successful valve implantation in horizontal aorta with TAVI was achievable; the rate of technical success was high. In this series, the Neo2 device showed a lower incidence of post-procedural gradients. The Neo2 short stent-frame and stabilizing arches may allow for greater coaxiality and stability, justifying this hemodynamic advantage.

Dr. Omar Tupayachi

Dr. Omar Tupayachi.
Member of the Editorial Board of SOLACI.org.

Original Title: Outcomes of transfemoral TAVR using two new‐generation devices in patients with horizontal aorta.

Reference: Eckel C, Kim WK, Wasif B, Grothusen C, Elsässer A, Dohmen G, Charitos EI, Sossalla S, Möllmann H, Blumenstein J. Outcomes of transfemoral TAVR using two new-generation devices in patients with horizontal aorta. Catheter Cardiovasc Interv. 2024 Jun 11. doi: 10.1002/ccd.31094. Epub ahead of print. PMID: 38860614.


Subscribe to our weekly newsletter

Get the latest scientific articles on interventional cardiology

More articles by this author

Long-Term Predictors of Valve Deterioration After TAVI

Transcatheter aortic valve implantation (TAVI) has become an innovative technology for the treatment of patients with severe aortic valve stenosis. Initially implemented in elderly...

Coronary access after TAVI with fourth- and fifth-generation Evolut valves: the EPROMPT-CA study

The expansion of transcatheter aortic valve implantation (TAVI) toward younger and lower-risk patients has increased the relevance of post-procedural coronary access. In this context,...

TEER plus optimal medical therapy versus medical therapy alone in functional mitral regurgitation

Mitral regurgitation (MR) is a highly prevalent valvular heart disease that, in advanced stages and when left untreated, is associated with reduced quality of...

VECTOR: First Percutaneous Aorto-Coronary Bypass Case, a New Conceptual Approach

Coronary obstruction represents one of the most severe complications associated with transcatheter aortic valve implantation, particularly in valve-in-valve scenarios involving surgical bioprostheses, narrow aortic...

LEAVE A REPLY

Please enter your comment!
Please enter your name here

Related Articles

SOLACI Sessionsspot_img

Recent Articles

Duration of Smoking Cessation and Risk of Amputation After Revascularization in Critical Limb Ischemia

Critical limb ischemia (CLI) is associated with high rates of amputation and mortality. Although smoking cessation improves outcomes after revascularization, the impact of the...

Conservative management of endoleaks in complex aortic endografts under CT angiography follow-up

Endoleaks remain one of the leading causes of reintervention after endovascular repair of complex aortic aneurysms using fenestrated and/or branched endografts (F/B-EVAR). Traditionally, type...

Long-Term Predictors of Valve Deterioration After TAVI

Transcatheter aortic valve implantation (TAVI) has become an innovative technology for the treatment of patients with severe aortic valve stenosis. Initially implemented in elderly...